BIOMET UK LTD. OXFORD UNI TIB BRG MED SZ4; PROSTHESIS, KNEE Back to Search Results
Catalog Number 154627
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 11/09/2023
Event Type  Injury  
Event Description
It was reported that the patient underwent revision surgery due to infection.Poly exchange and washout were performed.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).G2-foreign-australia d10.Item#: 154611;lot#:856397 ;item name: oxford uni tib tray 47x30 rllm; item#: 154601;lot#:861147 ;item name: oxford uni femoral md; item#: 6198b001;lot#:antibiotic simplex ;item name: bll058; investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.During the investigation process a review of the sterile certifications were not reviewed.All devices manufactured follow acceptable sterilization processes according to published iso/aami/astm & eu guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.All products manufactured follow appropriate standards, and the reported infection occurred above ninety days; therefore, implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD UNI TIB BRG MED SZ4
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18282944
MDR Text Key329943812
Report Number3002806535-2023-00416
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2010
Device Catalogue Number154627
Device Lot Number865238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received12/07/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/06/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-