• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Device Problems Fracture (1260); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
A2: age at time of event: 18 years or older.
 
Event Description
It was reported that stent fracture occurred.Vascular access was gained via contralateral approach.The 100% stenosed target lesion was located in the severely calcified and moderately tortuous superficial femoral artery (sfa).A 6x150, 130 cm eluvia drug-eluting vascular stent system was selected for use.The lesion was predilated with a peripheral angioplasty balloon.Difficulty occurred crossing the lesion where the chronic total occlusion (cto) was and the stent delivery system would not advance.During deployment, 20-30mm of the stent was deployed and broke off from the rest of the stent system.The 120-130mm of the broken part of the stent became stuck in the sheath.The sheath was then removed and replaced.The remaining 20-30mm portion of the stent was implanted.There were no patient complications, and the patient status was good post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18283275
MDR Text Key329947218
Report Number2124215-2023-69168
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0031205587
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE SHEATH: 6X45 TERUMO DESTINATION.; GUIDEWIRE: .035" GWA.
-
-