Model Number 71992-01 |
Device Problem
Device Alarm System (1012)
|
Patient Problems
Hypoglycemia (1912); Nausea (1970); Diaphoresis (2452); Confusion/ Disorientation (2553)
|
Event Date 11/29/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An alarm issue was reported with the adc device.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced sweating, hiccups, dizziness, tiredness, nausea and was unable to self-treat.The customer received sugar and water by a third party for treatment.There was no report of death or permanent impairment associated with this event.
|
|
Event Description
|
An alarm issue was reported with the adc device in use with huawei, os version 9 and software version 2.10.1.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced sweating, hiccups, dizziness, tiredness, nausea and was unable to self-treat.The customer received sugar and water by a third party for treatment.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug was properly seated, and no physical damage was observed on the sensor patch.The sensor was found to be in state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.The reported sensor was activated with a retained reader and linearity testing was performed, all results were within specification.Both high and low glucose alarms were successfully activated.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
|
Search Alerts/Recalls
|
|