• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG Back to Search Results
Model Number 6662
Device Problem Battery Problem (2885)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388); Movement Disorder (4412)
Event Date 11/10/2023
Event Type  Injury  
Manufacturer Narrative
Date of event estimated.
 
Event Description
It was reported the patient's ipg was dead, and it is unknown if the device depleted prematurely.The patient was not receiving therapy, and the ipg was deemed inoperable.Surgical intervention took place where the ipg was explanted and replaced to address the issue.Effective stimulation was restored.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18285832
MDR Text Key329995165
Report Number1627487-2023-05814
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/10/2023
Device Model Number6662
Device Lot Number8166194
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DBS BURR HOLE CAP; DBS BURR HOLE CAP (X2); DBS EXTENSIONS (X2); DBS LEAD; DBS LEAD
Patient Outcome(s) Other;
Patient SexMale
Patient Weight73 KG
-
-