Model Number N/A |
Device Problems
Battery Problem (2885); Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to surgery on 06 nov 2023, pulsavac had debris in the sterile packaging, see attached photo.There was no patient involvement, impact, or harm.Due diligence information complete.No additional information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).E1 telephone number: (b)(6).G2 foreign: finland.Evaluation and investigation are in process.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).E1: telephone number: (b)(6).No product was returned; functional, and/or dimensional evaluations could not be performed.Visual evaluation of the provided picture found black debris throughout the product sterile packaging.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|