MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM Back to Search Results
Model Number N/A
Device Problems Battery Problem (2885); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Event Description
It was reported that prior to surgery on 06 nov 2023, pulsavac had debris in the sterile packaging, see attached photo.There was no patient involvement, impact, or harm.Due diligence information complete.No additional information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).E1 telephone number: (b)(6).G2 foreign: finland.Evaluation and investigation are in process.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).E1: telephone number: (b)(6).No product was returned; functional, and/or dimensional evaluations could not be performed.Visual evaluation of the provided picture found black debris throughout the product sterile packaging.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18287209
MDR Text Key329975206
Report Number0001526350-2023-01622
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00515047500
Device Lot Number70933348
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-