Model Number 25052
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Event Description
The customer reported the radical-7 turns on and off during use.There was no patient impact or consequence reported.
 
Manufacturer Narrative
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Manufacturer Narrative
Additional manufacturing narrative: other, other text: the returned device was evaluated.There was no damage or defects observed during visual inspection.The device could be powered on and off using ac and dc power.Alarm conditions were audible and visual.The reported power issue was not observed during lab testing.
 
Event Description
The customer reported the radical-7 turns on and off during use.There was no patient impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7 HANDHELD
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
tahereh sedighi
15750 alton pkwy
irvine, CA 92618
9492977862
MDR Report Key18287739
MDR Text Key330019727
Report Number3019388613-2023-00276
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K110028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25052
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-