Model Number 1458Q/86
Device Problems Therapy Delivered to Incorrect Body Area (1508); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem Discomfort (2330)
Event Date 11/20/2023
Event Type  Injury  
Event Description
It was reported that the patient presented for an implant procedure.After the procedure, it was noted that the left ventricular lead exhibited phrenic nerve stimulation.It was noted the lead had dislodged.The physician attempted to reposition the lead but the stylet failed to advance in the lead.The lead was explanted and replaced.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18288080
MDR Text Key329993564
Report Number2017865-2023-94231
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000140691
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/20/2023
Initial Date FDA Received12/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/52 TENDRIL STSLEADS; 2088TC/58 TENDRILLEADS; PM3562 QUADRA ALLURE
Patient Outcome(s) Required Intervention;
Patient Age85 YR
Patient SexFemale
Patient Weight67 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-