The following has been reported during training by a fresenius kabi trainer: "pumps are alarming secondary occlusion alarm when trying syringes (medline 20 ml syringes syr12001) or secondary lines (bbraun secondary set v1921 40 in).I have swapped out 10 pumps and 10 sets and i tried different syringes and different secondary set, all have the same occlusion alarm.I have tried taking the set off and putting it back on, i have tried pulling the syringe barrel out and back to break the stiction.When i used my set i brought from home with a different lot- mine worked.The secondary set and the syringe both worked on my set that i brought with me." reporting due to the referenced issue.Occurred during training; no adverse effects were reported.More information is needed to complete the investigation.
|