BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
|
Back to Search Results |
|
Model Number L331 |
Device Problems
Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the day following the implant procedure of this implantable device, a surgical revision procedure was performed.During this procedure, it was noted that one of the device's set screws was damaged and fell out of the seal plug.The physician subsequently explanted and replaced this device to resolve the event and no additional adverse patient effects were reported.This device is expected to be returned for analysis, but has not yet been received.
|
|
Event Description
|
It was reported that the day following the implant procedure of this implantable device, a surgical revision procedure was performed due to worsening electrical parameters.The specific unsuitable electrical measurements were not specified.During this procedure, it was noted that one of the device's set screws was damaged and fell out of the seal plug.The physician subsequently explanted and replaced this device to resolve the event and no additional adverse patient effects were reported.This device is expected to be returned for analysis, but has not yet been received.
|
|