Model Number 1111178
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2020
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator unexpectedly shutdown and randomly would power on by itself.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer¿s complaint was not confirmed.However, the device failed during run in.The devices board needs replaced to address the issue.Additionally, the right side panel, blower, box mount, flow tube, pressure tube and manifold were replaced for contamination which was not related to the malfunction/issue.The board was unable to be replaced due to a software issue.The device was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A30
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18288792
MDR Text Key330012774
Report Number2518422-2023-34244
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113053
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1111178
Device Catalogue Number1111178
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2020
Initial Date Manufacturer Received 06/15/2020
Initial Date FDA Received12/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A