SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Smoking (1585); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found the reported malfunction could not be confirmed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported that the high flow insufflation unit could not insufflate during a tlh (total laparoscopic hysterectomy).The equipment was not set correctly, and the surgeons had a hard time visualizing the operation field due to too much smoke and was challenging to conduct the operation.However, the procedure was completed after a delay of 30 minutes while the patient was under general anesthesia.The whole operating room and hallway smelled of polluting smoke, there were no reports that any emergency/urgent issues/adverse conditions arose that required medical intervention or treatment.There is no evidence that a life-threatening event occurred, or that the patient or heath care staff developed a permanent disability or permanent damage that caused substantial disruption in their ability to conduct normal life functions, or that additional intervention was required to preclude permanent impairment of a body function or damage to a body structure.Though there was a delay this delay did not cause harm/injury/adverse clinical impact.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18289103
MDR Text Key330253702
Report Number3002808148-2023-13944
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received02/14/2024
Supplement Dates FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-