Model Number CDI510H
Device Problem Calibration Problem (2890)
Patient Problem Insufficient Information (4580)
Event Date 05/23/2023
Event Type  malfunction  
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, they noticed a calibration slope error on: arterial pco2, po2, ph.It is unknown if there was a delay in the procedure, if surgery was completed successfully, and if the product malfunction cause or contribute to an injury, and whether there was a blood loss.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.*no patient involvement.*the product was changed out.
 
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on december 8, 2023.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date) d9 (updated device availability) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h3 (updated device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 11, 3331, 4114, 3221, 4315) the complaint sample was not returned; therefore, a thorough investigation could not be performed, and a definitive root cause could not be determined.A retention sample from the same product code and lot number was set up for gas calibration with the cdi 500 and cdi 540.The retention sample was found to be successfully gas calibrated with no issues.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key18290429
MDR Text Key330039509
Report Number1124841-2023-00276
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2023
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot Number3A16K
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/15/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received01/03/2024
Supplement Dates FDA Received01/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-