ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problems Difficult to Open or Close (2921); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 12/8/2023.D4: batch # a9dn0g.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 4 conforming clips. upon testing, the jaws open and close without any difficulties.In addition, the device locked out as intended.  although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: do not excessively twist or torque the instrument jaws when positioning or firing the instrument on a tubular structure or vessel.Excessive twisting or torquing may result in clip malformation.Do not insert the clip applier through a trocar if a clip is present in the jaws.This may result in clip malformation, dislodged clips, or damage to the instrument.If a clip is present in the jaws, fully squeeze the trigger against the handle, then fully release the trigger to release the clip from the jaws before inserting the device through the trocar.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the clip could not be fed into the jaw, so the surgeon grasped the trigger forcibly, but the jaws became not to open and could not be used.The device was used less than 10 firings.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18292010
MDR Text Key330054604
Report Number3005075853-2023-09191
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberA9DN0G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2023
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-