• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Catalog Number NGE-017115
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2023
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1: customer (person) phone = (b)(6).G4: pma/510(k) number = exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
As reported, during ureteroscopic laser lithotripsy, the user detected that the wire of an ngage nitinol stone extractor detached while removing stones.Another identical device was utilized to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.
 
Manufacturer Narrative
Correction: h6 (annex g).Investigation ¿ evaluation: as reported, during ureteroscopic laser lithotripsy, the user detected that the wire of an ngage nitinol stone extractor detached while removing stones.Another identical device was utilized to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.A document-based investigation was performed including a review of complaint history, device history record (dhr), instructions for use (ifu), manufacturing information (mi), and quality control procedures, as well as a visual inspection of the device were conducted.One ngage nitinol stone extractor was returned in an open package with label.Wires from the basket formation pulled free from the basket sheath.The shrink tube and glue that secures the basket to the basket sheath had not held the two components together.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no confirmed recorded non-conformances related to the failure mode.A lot history search found 5 complaints had been reported for this lot.All complaints were for the same issue.The information provided upon review of complaint file, device history record, complaint history, and quality control documents did not provide evidence to support that the device was manufactured out of specification or to suggest items in the lot or similar devices in the field or in house were nonconforming.Cook also reviewed product labeling.The product ifu, t_shef_rev1; the ifu did not provide any information related to the reported issue.Based on the information provided, inspection of the returned device, and the results of the investigation, a cause has not been determined.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18295448
MDR Text Key330829102
Report Number1820334-2023-01667
Device Sequence Number1
Product Code FFL
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNGE-017115
Device Lot Number15446255
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/17/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received04/01/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-