• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problems No Device Output (1435); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  malfunction  
Manufacturer Narrative
The test strips' expiration date was not provided.The coaguchek xs meter serial number was not provided.The test strips were requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling: coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.
 
Event Description
We received an allegation of questionable inr results for 1 patient tested with his coaguchek xs pst meter compared to a different coaguchek xs meter at the doctor's office and to an unknown laboratory method.On an unknown date and time, the patient had a meter result of 2.5-3.0 inr.On (b)(6) 2023 at 2:28 pm, the doctor's office meter gave a result of >8.0 inr and at 2:52 pm the laboratory result was 7.1 inr.On (b)(6) 2023 at 3:48 pm, the doctor's office meter gave a result of >8.0 inr and at 4:02 pm the laboratory result was 5.5 inr.The patient's therapeutic range was 2.5-3.5 inr.
 
Manufacturer Narrative
Section h6, medical device problem code, and type of investigation code were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18296327
MDR Text Key330830616
Report Number1823260-2023-03915
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702127104
UDI-Public00365702127104
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04625374160
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/15/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received01/09/2024
Supplement Dates FDA Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CREON; FENTANYL PATCH; GENERIC NORFLEX; LEXAPRO 20MG; OMEPRAZOLE; OXYCODONE ACETAMINOPHEN; WARFARIN 4MG
Patient Age35 YR
Patient SexMale
-
-