Catalog Number 04625315160
Device Problems Misassembly by Users (3133); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/30/2023
Event Type  malfunction  
Event Description
We received an allegation of questionable inr results for 1 patient tested on a coaguchek xs meter compared to another coaguchek xs meter at a different facility (coagulation clinic).At 10:25 am the patient was tested in the pain management clinic and received a meter result of 1.0 inr.The patient was then tested in a coagulation clinic in the same building and received a meter result of 1.9 inr.The time difference between the test measurements was within minutes.The physician believed that 1.9 inr was the correct result.The patient¿s therapeutic range was not provided.Reportedly, the result of 1.9 inr was alleged to be outside the patient's therapeutic range.
 
Manufacturer Narrative
The coaguchek xs meter serial number used at the pain management clinic was (b)(6).The reporter stated that the meter was cleaned due to some minor contamination.The test strips were requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Manufacturer Narrative
The type of investigation code was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18296331
MDR Text Key330142558
Report Number1823260-2023-03914
Device Sequence Number1
Product Code GJS
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number04625315160
Device Lot Number62215312
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received01/09/2024
Supplement Dates FDA Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
WARFARIN
Patient SexFemale
-
-