Model Number 101-9812
Device Problem Migration (4003)
Patient Problem Pain (1994)
Event Date 10/01/2023
Event Type  Injury  
Manufacturer Narrative
Block b3: exact date unknown, event occurred in (b)(6) 2023.
 
Event Description
It was reported the patient experienced a return of their pre-implant pain following several non-device related falls.As a result of the falls, the indirect decompression spacer migrated off of the spinous process.The patient underwent a revision procedure to replace the migrated spacer.
 
Event Description
It was reported the patient experienced a return of their pre-implant pain following several non-device related falls.As a result of the falls, the indirect decompression spacer migrated off of the spinous process.The patient underwent a revision procedure to replace the migrated spacer.
 
Manufacturer Narrative
Analysis of the returned device exhibited normal characteristics and revealed no anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Type of Device
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18298346
MDR Text Key330111342
Report Number3006630150-2023-07739
Device Sequence Number1
Product Code NQO
UDI-Device Identifier00884662000543
UDI-Public00884662000543
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number101-9812
Device Catalogue Number101-9812
Device Lot Number28282296
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received12/09/2023
Supplement Dates Manufacturer Received01/26/2024
Supplement Dates FDA Received01/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-