• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 490102
Device Problem Device Alarm System (1012)
Patient Problems Low Blood Pressure/ Hypotension (1914); Tachycardia (2095); Increased Respiratory Rate (2486)
Event Type  Death  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As per medwatch mw5148160: "patient had laparoscopic surgery with small intestine resection on (b)(6) 2023.On (b)(6) 2023 at 2355 patient became hypotensive (bp 75/50), tachycardic (hr 20's), and tachypneic (rr high-40's).Rapid response called and the patient was transferred to the icu.Utilizing the bbraun infusomat space pumps the patient was started on levophed, and then vasopressin and neosynephrine are added.The bbraun administration sets utilized (infusomat space 15 drip/ml 120 inch tubing) are the subject of a bbraun urgent medical device correction notice dated october 23, 2023 and this notification was received via us mail at the hospital on november 6, 2023.There are no unaffected tubing administration lots in commercial circulation at this time with new improved inventory due on november 30, 2023 per the bbraun device correction notice.As noted in the correction notice bbraun has identified that their outer diameter tubing may vary in size leading to increased air alarms, including false air alarms as well as the creation of microbubbles in the tubing administration sets, further leading to additional air alarms.Due to multiple air alarms generated by the bbraun infusomat space pumps in use (8 multiple bbraun infusomat space pumps running for this critically ill patient) as a result of the administration set issue this patient experienced further instability leading to the interruption of the infusion of vasopressors, levophed and neosynephrine in a hemodynamically unstable patient.The patient has subsequently passed away on (b)(6) 2023.As such we are reporting this event to the fda.Reference reports: mw5148158, mw5148159.".
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A review of manufacturing records was performed and indicated that there were no quality issues during the manufacturing of this lot related to the reported issue.No sample was provided for evaluation; however, b.Braun has initiated a voluntary medical device correction 2523676-9/26/23-004-c for multiple batches of infusomat® pump sets manufactured bewveen 01 january 2022 - 17 august 2023.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT®
Type of Device
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key18300558
MDR Text Key330162605
Report Number2523676-2023-00819
Device Sequence Number1
Product Code FPA
UDI-Device Identifier04046964186127
UDI-Public(01)04046964186127
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number490102
Device Lot Number0061896875
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2023
Initial Date FDA Received12/11/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death; Life Threatening;
-
-