Catalog Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem Brain Injury (2219)
Event Date 11/29/2023
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported a perforator (id 261221) did not stop after drilling through the bone (first hole) causing perforation of the dura with minimal blood loss.The dura was repaired: "approximated and spongostan laid over".The procedure was completed with a replacement product available with no significant surgical delay.The manufacturer of the drill used with the perforator was an electronic medtronic (midas rex mr8).The perforator clicked in place with the drill and the recommended spring tests was performed between each burr hole.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.Perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the returned unit was found to work as intended, and met all acceptance criteria.The complaint could not be verified through failure analysis.Root cause - the root cause is undetermined and was unable to be confirmed in the complaint evaluation.Although the complaint was not confirmed, root cause was investigated using the risk documentation.Per the risk document, potential causes of failure include: user misuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key18302201
MDR Text Key330178321
Report Number3014334038-2023-00188
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot Number7230472
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/11/2023
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age63 YR
-
-