SYNTHES GMBH AIMING ARM/ RADIOLUCENT; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 03.043.029
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2023
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h6: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the device 03.043.029, aiming arm/ radiolucent had both of the hooks of the latch broken.The device shows the signs of usage and broken hooks of latch and no evidence of hammer marks were observed, therefore, a potential cause cannot be established with the provided information.The semi-extended parapatellar approach surgical technique guide se_814687 ad was reviewed.Following relevant statements were found.Precautions: -ensure that the connection between the nail and the insertion handle is tight.Retighten if necessary, after hammering and prior to the attachment of the aiming arm.-do not attach the aiming arm to the insertion handle at this point.-do not exert forces on the aiming arm.These forces may prevent breakage of the device.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device 03.043.029, aiming arm/ radiolucent would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2023, the tna tibial nail advanced was inserted into the tibia with mallet adapter and aiming jig attached.Upon insertion the carbon fiber latches on the aiming arm broke off.The surgery was successfully completed with 2min of surgical delay.Fragments were generated and easily removed.A back-up instrument set was used.There was no patient consequence was reported.This report involves one (1) aiming arm/ radiolucent.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM/ RADIOLUCENT
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18302215
MDR Text Key330178490
Report Number8030965-2023-15556
Device Sequence Number1
Product Code JDS
UDI-Device Identifier10886982297396
UDI-Public10886982297396
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201336
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.043.029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - NAILS: TIBIAL.
-
-