Catalog Number ENCR402300 |
Device Problem
Premature Activation (1484)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone: (b)(6).Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 8484768.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
Event Description
|
The healthcare professional reported that during a stent-assisted endovascular embolization procedure, just after the complaint device, a 4mm x 23mm no tip enterprise¿ 2 vascular reconstruction device (encr402300 / 8484768), ¿the stent was found released automatically and the stent body was prematurely separated from the delivery wire.¿ the physician removed the stent and switched to a new stent to complete the procedure using the original concomitant microcatheter (unspecified brand).There was no report of any negative patient impact.On 11-dec-2023, the cerenovus sales representative reported that additional information is not available.
|
|
Manufacturer Narrative
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that multiple attempts to obtain product for analysis were unsuccessful.If product is returned or information provided at a later date, the file will be updated accordingly.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 8484768.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.With the information available and without the complaint product available to be returned for analysis, the reported issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation/interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
Search Alerts/Recalls
|
|