MAUDE Adverse Event Report: RESPIRONICS, INC. DS2 ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DS2 ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX520H11C
Device Problems Mechanical Problem (1384); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2023
Event Type  malfunction  
Event Description
The manufacturer received information regarding a dreamstation 2 cpap device.The patient states the button on their device intermittently stops working.The patient also state that the device got hot too the touch and almost burned her.The device was returned to a third party service center for evaluation.During the evaluation of the device, it was determined that the pca has functional failure.The device was scrapped per deviation v01500886.There was no allegation of serious or permanent harm or injury.No medical intervention was required for the patient.The third party service center scrapped the device.At this time, no further investigation can be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DS2 ADV AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18304064
MDR Text Key330191198
Report Number2518422-2023-34748
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX520H11C
Device Catalogue NumberDSX520H11C
Was Device Available for Evaluation? No
Date Manufacturer Received04/12/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/17/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-