• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTOSCOPE SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTOSCOPE SHEATH Back to Search Results
Model Number A22041A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: the insulation insert is broken off.No other issues were found at evaluation.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the resection sheath, 24 fr.Had a dent in the shaft.There were no reports of patient harm.The device was returned for evaluation and during the evaluation the following reportable malfunction was found: the insulation insert is broken off.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 24 FR.
Type of Device
RESECTOSCOPE SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18307672
MDR Text Key330561643
Report Number9610773-2023-03599
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020961
UDI-Public04042761020961
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22041A
Device Lot Number175W-0122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-