• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH RPD150; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH RPD150; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number 10-90-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the essenza pump.The incident occurred in the netherland.Pump can be used since this error is resolved.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that, during a prcedure, the essenza rpd150 pump displayed error indicating deviation of the observed direction of rotation from the set direction of rotation (e2350) and stopped.After turning the knob to zero (0) and back to initial flow the pump started again.Procedure could be finished.There is no report of any patient injury.
 
Manufacturer Narrative
The involved pump is a left vent pump connected to cu.Log provided and found to be empty.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RPD150
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18309614
MDR Text Key330262610
Report Number9611109-2023-00634
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier04033817902997
UDI-Public010403381790299711230220
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K232291
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10-90-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/23/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/20/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-