Catalog Number D134805
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Cardiac Tamponade (2226)
Event Date 11/13/2023
Event Type  Injury  
Event Description
It was reported a patient underwent an atrial fibrillation (afib) cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and experienced cardiac tamponade treated with a pericardiocentesis.Transseptal puncture performed, and ablation performed of the left pulmonary vein.Anesthesia notified the physician that the patients' blood pressure dropped.A high force reading was displayed on the carto 3 system and a pericardial effusion was diagnosed via intracardiac echo after the patient's pressure drop was observed.A pericardiocentesis was performed by the physician and 900 cc of fluid was removed from the pericardial space.Protamine was reversed.The patient was stable and transferred to ccu (critical care unit) for observation.The physician¿s opinion on the cause was that it was patient condition related.Patient has fully recovered.
 
Manufacturer Narrative
The device was returned to biosense webster (bwi) for evaluation.A visual inspection and revision of all features were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.The device features were reviewed, and no issues were observed during the product investigation.A manufacturing record evaluation was performed for the finished device 31151927l number, and no internal action related to the complaint was found during the review.No malfunction was observed during the product analysis.The root cause of the adverse event remains unknown.There may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.The physician¿s opinion on the cause of this adverse event was the patient condition.The instruction for use (ifu) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation or tamponade.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18311053
MDR Text Key330285630
Report Number2029046-2023-02914
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31151927L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BAYLIS TRANSEPTAL NEEDLE.; NGEN CONSOLE.; STERILMED CS CATHETER (REPROCESSED).; STERILMED ULTRASOUND CATHETER (REPROCESSED).; UNK_CARTO 3.; UNK_CARTO VIZIGO SHEATH.; UNK_OCTARAY NAV.
Patient Outcome(s) Required Intervention; Life Threatening; Hospitalization;
Patient SexMale
-
-