It was reported that during an arthroscopy, the capturing plate on the tip (on the upper jaw) of the firstpass suture passer was detached and fall into the patient's wound, it was removed with graspers.The procedure was completed using a back-up device.There was no surgical delay and no further complications were reported.
|
H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was not returned in any original packaging.The suture capture is missing from the upper jaw and was not returned.There are no other visible defects.A functional evaluation showed pulling the lever will close and lock the jaw.Pulling the lever and trigger simultaneously deploys the suture passer needle.These failures have been determined to be related to the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|