• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM CARY WAVE CLINICAL PLATFORM; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM CARY WAVE CLINICAL PLATFORM; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the wave configuration was not updated.The wave clinical platform is software intended to route, store, and display data, alarms, results and diagnostic information from medical devices, electronic medical records (emr), and clinical information systems (cis).The wave clinical platform is a remote monitoring platform that displays physiologic data, waveforms, alarms, results and diagnostic information routed through the platform from supported devices and systems.The wave clinical platform is intended for use in hospital or hospital type environments.The wave clinical platform is intended to be used by healthcare professionals for the following purposes: to remotely consult regarding patients¿ statuses; to remotely review other standard or critical near real-time patient data, waveforms, alarms, and results in order to utilize this information to aid in clinical decisions.The user manual contains warnings such as "the wave clinical platform is intended to supplement and not replace any part of the hospital's device monitoring or electronic data management systems.Do not rely on the wave clinical platform product as the sole source of alarms".The hrc technician arranged to get the rooms configured for the customer to resolve the issue.
 
Event Description
Hillrom received a report from a hillrom technician stating the code calls were not arriving to voalte phones.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVE CLINICAL PLATFORM
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
HILL-ROM CARY
1225 crescent green dr., suite 300
cary NC 27518
Manufacturer Contact
estephania cardenas
1225 crescent green dr., suite 300
cary, NC 27518
8129310130
MDR Report Key18313537
MDR Text Key330940855
Report Number2027454-2023-00097
Device Sequence Number1
Product Code MWI
UDI-Device Identifier00856128007007
UDI-Public00856128007007
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/07/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-