Event Description
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to non function and occlusion.First, it was reported that the ra lead was extracted without issue.Then, a spectranetics lld ez lead locking device (lld ez) was inserted into the rv to provide traction.While using a spectranetics 16f glidelight laser sheath and a spectranetics visisheath dilator sheath to remove the rv lead, the lead tip freed.However, the patient''s blood pressure dropped, with a pericardial effusion detected and growing.Rescue efforts began, including rescue balloon and sternotomy.A small rv perforation was discovered and repaired, and the patient survived the procedure.This report captures the lld ez providing traction within the rv lead when the perforation occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
 
Manufacturer Narrative
A4): patient''s weight unk.B7): other relevant history unk.H3): the device was discarded, thus no investigation could be completed.H6): cardiac perforation is a known risk of complication with use of the lld.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LLD EZ LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key18314623
MDR Text Key330328220
Report Number3007284006-2023-00088
Device Sequence Number1
Product Code DRB
UDI-Device Identifier20813132023076
UDI-Public(01)20813132023076(17)250227(10)FLP23A25A
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberFLP23A25A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC 5076 RA PACING LEAD; MEDTRONIC 6935 RV ICD LEAD; SPECTRANETICS 16F GLIDELIGHT LASER SHEATH; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS VISISHEATH DILATOR SHEATH
Patient Outcome(s) Required Intervention; Life Threatening;
Patient Age46 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-