• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134805
Device Problems Material Puncture/Hole (1504); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2023
Event Type  malfunction  
Manufacturer Narrative
The product investigation was completed.Device evaluation details: the smart touch bidirectional sf device was returned to biosense webster (bwi) for evaluation.A visual inspection and magnetic sensor functionality test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed a hole and reddish material in the pebax.The device was connected to the carto 3 system and it was recognized; however, error 105 was displayed on the screen due to an open circuit in the tip area.The blood found inside the pebax area may contribute to the magnetic issue.The root cause of the pebax damaged could be related to the procedure however this cannot be conclusively determined.A manufacturing record evaluation was performed for the finished device 31049481l number, and no internal action was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number:(b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, the carto 3 system displayed a magnetic sensor error.The smarttouch sf catheter connections were re-seated without resolution.The cable was replaced without resolution.The smarttouch sf catheter was replaced and the issue resolved.The procedure continued.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18315255
MDR Text Key330354701
Report Number2029046-2023-02930
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31049481L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2023
Date Manufacturer Received11/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM; VIZIGO SHEATH
-
-