Catalog Number 80224 |
Device Problems
Break (1069); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 01/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a recanalization procedure in the left iliac artery via the femoral artery approach, the distal end of the catheter allegedly had a leak.There was no reported patient injury.
|
|
Event Description
|
It was reported that during a recanalization procedure in the left iliac artery via the left femoral artery approach, the distal end of the catheter allegedly had a leak.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation and a physical investigation was performed for the catheter.During physical investigation the tube of the rotarex was totally broken at 1 cm from the tip of the catheter.The catheter was delivered without collecting bag connection tube.Reported leak was not observed and was not confirmed during physical investigation.Therefore, the investigation is unconfirmed for the reported leak issue.However, the investigation is confirmed for the identified break issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: (expiration date: 01/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|