• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problems Failure to Fold (1255); Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported with a description of lenses defective.Additional information was requested.
 
Event Description
Additional information was received and physician stated that the lenses didn't fold well and he wasn't comfortable implanting them.
 
Manufacturer Narrative
Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.The product has not been received to evaluate.It is unknown if a qualified viscoelastic was used.The ifu ( instructions for use) instructs: during device preparation and implantation of the company iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Company account manager indicated walking the surgeon through how to trouble shoot the issue next time it was encounter.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18317959
MDR Text Key330432912
Report Number1119421-2023-02158
Device Sequence Number1
Product Code KYB
UDI-Device Identifier00380652394789
UDI-Public00380652394789
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAU00T0
Device Lot Number15599714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2023
Initial Date FDA Received12/13/2023
Supplement Dates Manufacturer Received01/16/2024
Supplement Dates FDA Received02/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-