• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, the guidewire had failed to advance in left ventricular (lv) lead.The lv lead was removed and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of ¿the guidewire could not move in and out smoothly¿ was confirmed.As complete lead was returned in one piece.Visual and x-ray examination of the lead found the inner coil bunching up at the connector region.The cause of the reported event was isolated to the inner coil damaged consistent with procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18320542
MDR Text Key330467170
Report Number2017865-2023-94531
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000136638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-