MEDTRONIC EUROPE SARL COBALT¿ XT HF CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTPA2D4 |
Device Problems
Mechanical Problem (1384); Under-Sensing (1661); Decreased Sensitivity (2534)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during the implant procedure the r waves were within normal limits when measured through the analyzer, but when the right ventricular (rv) lead was plugged into the cardiac resynchronization therapy defibrillator (crt-d), the r waves were very low and undersensing was noted.The lead was unplugged and checked again via analyzer with normal measurements but again when plugged into the crt-d the measurement was low.It was suggested that there may be an issue with the rv lead connector pin and the physician indicated that there were problems when putting the stylet though the lead.Additionally, the physician thought there may be an issue with the crt-d header.The crt-d was removed and replaced; however, when the rv lead was plugged into the new crt-d the measurements were still low.The lead was removed and replaced which resulted in normal measurements.No patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|