• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC MYOSURE REACH TISSUE REMOVAL DEVICE; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC MYOSURE REACH TISSUE REMOVAL DEVICE; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 10-401FC
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2023
Event Type  malfunction  
Event Description
It was reported that during a myosure procedure on (b)(6) 2023, during the procedure the physician reported that the console screen displayed error ¨excessive torque¨, later the physician reported that a piece of the blade broke off and was floating in the patient´s uterus.The physician immediately removed the device and used graspers to retrieve the broken piece.The staff saved the broken piece for further testing and the procedure was aborted.No harm to the patient reported.No additional information available.
 
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE REACH TISSUE REMOVAL DEVICE
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
MDR Report Key18326585
MDR Text Key330498878
Report Number1222780-2023-00464
Device Sequence Number1
Product Code HIH
UDI-Device Identifier15420045504530
UDI-Public(01)15420045504530(17)260902(10)23J18R
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number10-401FC
Device Catalogue Number10-401FC
Device Lot Number23J18R
Was Device Available for Evaluation? No
Date Manufacturer Received11/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient SexFemale
-
-