An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device was loaded into a test 10f loader, deployed and retracted without difficulty or deformation, under non-physiological conditions.Additionally, an image was received from the field to show the device deformity.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.A 12f sheath was used to deliver the device, which could have contributed to the deformed deployment noted, as use of a larger sheath may influence deformations of the device.
|