Catalog Number 306593
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2023
Event Type  malfunction  
Manufacturer Narrative
Pr (b)(4): initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event Description
The patient was admitted to the hospital for renal insufficiency, and on (b)(6), 2023, the nurse-in-charge closed the patient's indwelling needle after administering intravenous fluids, and when using a prefilled catheter flosser, he found that the cartridge rod was disconnected from the piston and was unusable, and he was immediately replaced with a different prefilled catheter flosser, without harm to the patient.
 
Manufacturer Narrative
(b)(4) follow up.A device history record review was completed by our quality engineer team for provided material number 306593 and lot number 2096905.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.
 
Event Description
No additional information received.The patient was admitted to the hospital for renal insufficiency, and on (b)(6) 2023, the nurse-in-charge closed the patient's indwelling needle after administering intravenous fluids, and when using a prefilled catheter flosser, he found that the cartridge rod was disconnected from the piston and was unusable, and he was immediately replaced with a different prefilled catheter flosser, without harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 3ML SALINE FILL CHINA SP
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BD MDS DCHU
75 north fairway drive
vernon hills IL 60061
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18329100
MDR Text Key330973047
Report Number1911916-2023-00892
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065937
UDI-Public(01)30382903065937
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number306593
Device Lot Number2096905
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial