• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE PROGRESSA FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE PROGRESSA FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P7500A000181
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2023
Event Type  malfunction  
Event Description
Baxter received a report from a baxter technician stating the bed had power cord damage with exposed copper wires.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
 
Manufacturer Narrative
The baxter technician found the power cord needed to be replaced.It is necessary for the progressa¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Examine the plug for damage.Make sure the plug is a one-piece molded plug assembly.If it is not, replace the plug cord assembly.Replace any plug cord assembly that shows: discoloration of the plug molding, any signs of cracking, or loose fit of the plug blade.Replace the power cord, if damaged.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRESSA FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
estephania cardenas
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key18331869
MDR Text Key330963561
Report Number1824206-2023-01402
Device Sequence Number1
Product Code FNL
UDI-Device Identifier00887761000100
UDI-Public010088776100010011180312
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP7500A000181
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-