Model Number N/A |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).G2 ¿ foreign ¿ singapore.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
Event Description
|
It was reported that there is visible crack line on aseptic battery housing lid.This event is related to a malfunction that could potentially lead to a sterility issue/serious injury.However, no patient harm or further outcome was reported.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional and/or corrected information.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Upon receipt, the lid hinge was broken and there was corrosion on the contacts.The unit was scrapped.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|