Catalog Number 09038086190 |
Device Problems
High Test Results (2457); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cobas pure e 402 analytical unit serial number was (b)(6).The qc and calibration results were within specifications.Therefore a general reagent issue could be excluded.The customer found a cap in the reagent carousel that appeared to be from the pretreatment reagent.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable vitamin d total g3 elecsys e2g results from the cobas pure e 402 analytical unit.The initial result was 120 ng/ml with a data flag.The automatic repeat with a 1:2 dilution was 16.6 ng/ml.The second repeat result was 19.2 ng/ml.
|
|
Manufacturer Narrative
|
The investigation did not identify a product problem.The cause of the event could not be determined.
|
|