• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL III; VITAMIN D TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL III; VITAMIN D TEST SYSTEM Back to Search Results
Catalog Number 09038086190
Device Problems High Test Results (2457); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2023
Event Type  malfunction  
Manufacturer Narrative
The cobas pure e 402 analytical unit serial number was (b)(6).The qc and calibration results were within specifications.Therefore a general reagent issue could be excluded.The customer found a cap in the reagent carousel that appeared to be from the pretreatment reagent.The investigation is ongoing.
 
Event Description
There was an allegation of questionable vitamin d total g3 elecsys e2g results from the cobas pure e 402 analytical unit.The initial result was 120 ng/ml with a data flag.The automatic repeat with a 1:2 dilution was 16.6 ng/ml.The second repeat result was 19.2 ng/ml.
 
Manufacturer Narrative
The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN D TOTAL III
Type of Device
VITAMIN D TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18334250
MDR Text Key330593637
Report Number1823260-2023-04023
Device Sequence Number1
Product Code MRG
UDI-Device Identifier07613336172650
UDI-Public07613336172650
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
K210901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09038086190
Device Lot Number74466101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-