• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORTRAK 2 ENTERAL ACCESS SYSTEM; DH CORTRAK (EAS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORTRAK 2 ENTERAL ACCESS SYSTEM; DH CORTRAK (EAS) Back to Search Results
Model Number 20-0950
Device Problem Malposition of Device (2616)
Patient Problems Death (1802); Perforation (2001)
Event Date 08/12/2023
Event Type  Injury  
Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 14-dec-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.H3 other text : device not returned.
 
Event Description
It was reported a "small bore nasogastric tube placement attempted using cortrak guidance.Placement was successful after 5 attempt(s).Patient tolerated procedure poorly.[computed tomography] ct abdomen/pelvis showing concern for gastric perforation." a "stat ctap [computed tomography of abdomen and pelvis] and chest with gastric wall" noted a perforation involving the gastric antrum and the ng [nasogastric] tube was passing through the perforation with the tip ultimately in the inferior to the right hepatic lobe.Additional information received 27-nov-2023 stated the ng tube was inserted in the patient's right nare.The device was initially inserted on 11-aug-2023 at 0136.Additional information received on 08-dec-2023, exploratory laparotomy was performed.The current status of the patient is deceased.No details were provided regarding the cause of death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTRAK 2 ENTERAL ACCESS SYSTEM
Type of Device
DH CORTRAK (EAS)
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
MINNETRONIX, INC.
1635 energy park dr.
saint paul MN 55108
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key18334319
MDR Text Key330594336
Report Number3011270181-2023-00140
Device Sequence Number1
Product Code KNT
UDI-Device Identifier00350770472010
UDI-Public00350770472010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K220588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number20-0950
Device Catalogue NumberN/A
Device Lot Number20070471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Death; Other;
Patient Age82 YR
Patient SexMale
Patient Weight79 KG
-
-