Device Problem
Device-Device Incompatibility (2919)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a procedure on (b)(6) 2023, the aiming guide in question was unable to be disengaged.When attempting to slide it off distally, it was blocked by the distal screw.The guide was removed by removing the distal screw, sliding the guide off, and reinserting the screw.The procedure was completed successfully with no adverse patient impact.There was an unknown surgical delay.No further information is available.This report is for an unk - plate: fns.This is report 1 of 2 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d1, d2, d3, d4, g4 ¿ 510k: this report is for an unknown plate: fns/unknown lot.Part and lot numbers are unknown; udi number is unknown.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was not returned to depuy synthes, however photos were provided for review.The x-ray investigation revealed that unk - plates: fns had no defect found.According to an image provided in the surgical technique "femoral neck system" there is no visual indication that can confirm the device unable to assemble/disassemble issue reported.X-rays provided do not show evidence of the guide hitting the distal screw.However it is important as mentioned in the surgical technique page 10 precaution regarding the position of the guide: monitor the position of the wire during insertion and confirm the final position using the image intensifier.Over inserting guide wires could lead to damage to vital organs.Replace wires if they are bent after insertion.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the unk - plates: fns would not contribute to the complained device issue.Based on the investigation findings it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|