ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Therapy Delivered to Incorrect Body Area (1508); Capturing Problem (2891)
Patient Problem Discomfort (2330)
Event Date 11/30/2023
Event Type  malfunction  
Event Description
It was reported that during the implant procedure, unsatisfactory threshold value was noted on left ventricular (lv) lead and there was diaphragm stimulation.Finally, blood entered to the lead and make it impossible to continue.The lead was replaced.The patient is in stable condition.
 
Manufacturer Narrative
A complete lead was returned in one piece.Visual inspection of the lead insulation found no anomalies.The distal region and tip of the lead had blood.X-ray examination of the entire lead found no anomalies.The s-curve hump height was measured to be within product specification.The reported event inadequate capture was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18335412
MDR Text Key330604341
Report Number2017865-2023-94553
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000137198
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-