It was reported that prior to the procedure no water could be sprayed out from the device; the power was not enough.There was no harm or injury to patient as there was no patient involvement.The date the event occurred in not known.The battery showed signs of leaking.Due diligence is complete; however, no further information was received.As no additional information has been received, we are unable to provide further information.
|
This event is recorded by zimmer biomet under (b)(4).The following sections have been updated: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, h10.The device was returned without the original batteries in the pack.Functional testing found the device did power on and function normally in both modes when connected to a lab battery pack.Visual inspection of the interior of the pack found evidence the batteries had leaked electrolyte inside of the pack.There did not appear to be damage to the wiring of the pack.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to manufacturing issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|