MEDTRONIC HEART VALVES DIVISION EVOLUT FX VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVOLUTFX-29 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id l-evolutfx-2329 (lot: 0011840701); product type: 0195-heart valves; implant date n/a; explant date n/a product id d-evolutfx-2329 (lot: unknown); product type: 0195-heart valves; implant date n/a; explant date n/a medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that prior to the implant of this transcatheter bioprosthetic valve, during the loading process, the hospital technologist was unable to "marry" the valve as it appeared to be off the backplate.The loader attempted to "roll through" and fully encapsulate the valve without marrying.The valve load was inspected under fluoroscopy and a line to the fourth node was observed.It was attempted to re-prep, but the valve was bent sideways after marrying with the second attempt.Upon trying to take the valve out of the "honeycomb" and re-prep again, the valve was stuck.A valve outflow crown fracture was noted.A new valve and compression loading system were opened and loaded successfully by the medtronic representative.No adverse patient effects were reported; however, a ten minute procedural delay occurred.
|
|
Search Alerts/Recalls
|
|
|