Catalog Number 2078175
Device Problems Peeled/Delaminated (1454); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Manufacturer Narrative
Product performance engineering reviewed the incident information; however, the product was not returned to abbott vascular for analysis.A review of the electronic lot history record, corrective action tracking system, and similar incident query was not performed because the lot number was not reported.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling; therefore, no product-related corrective action will be implemented in this case.
 
Event Description
It was reported that the procedure was to treat a severely stenotic, heavily calcified and moderately tortuous lesion in the left superior femoral artery (sfa).The ht 14 command guide wire and a non-abbott crossing catheter met resistance with the anatomy during advancement.Despite no resistance being felt on removal of the guide wire, the black polymer coating was observed to be peeling at the distal end after it was removed.The physician confirmed nothing remained in the patient and the procedure was completed with a csi atherectomy device and a viper wire.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE COMMAND 14 GUIDE WIRE
Type of Device
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18337344
MDR Text Key330677207
Report Number2024168-2023-14085
Device Sequence Number1
Product Code DQX
UDI-Device Identifier08717648176685
UDI-Public08717648176685
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2078175
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2023
Initial Date FDA Received12/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-