ST. JUDE MEDICAL TACTIFLEX¿ ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number A-TFSE-DF |
Device Problem
Signal Artifact/Noise (1036)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2023 |
Event Type
malfunction
|
Event Description
|
During the atrial fibrillation/atrial tachycardia procedure, noise issues resulted in a procedural delay.The catheter was replaced and the procedure was completed with no adverse consequences to the patient.There was no problem with the electric potential when the catheter was inserted into the heart and connected to the tactisys and ensite, but after mapping, noise was noted on abl1-2 when rf delivery was started.The input box of the recording system was reconnected, the catheter was reconnected to the tactisys and ensite, the ensite and recording system were restarted, and it was re-grounded, with no resolution.The catheter was replaced which resolved the issue and the procedure was completed with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, h2, h3, h6 one bi-directional, curve d-f, tactiflex ablation catheter, sensor enabled was received for evaluation.Electrode 2 read as an open circuit during electrical testing.Further investigation revealed that conductor wire 2 was fractured proximal to the weld joint between the wire and the electrode ring, consistent with the detected open circuit, and the reported event.In addition, electrode 1 met specifications during electrical testing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of fractured conductor wire remains unknown.
|
|
Search Alerts/Recalls
|
|
|