This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The customer's complaint was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to thermal and mechanical induced impact, wear and tear, improper handling, mechanical overload like fall, shock or similar stress.The event can be prevented by following the instructions for use (ifu) which state: "before use warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center." olympus will continue to monitor field performance for this device.
|