MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG LT SM SIZE 4 PMA; OXFORD HXLPE BEARINGS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG LT SM SIZE 4 PMA; OXFORD HXLPE BEARINGS Back to Search Results
Model Number N/A
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Cyst(s) (1800); Failure of Implant (1924)
Event Date 11/26/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10 ¿ medical devices: oxf uni tib tray sz d lm pma; item# 154724; lot# 65676852.Oxf twin-peg cmntd fem sm pma; item# 161468; lot# 410880.Multiple mdr reports were filed for this event, please see associated reports: 3002806535 - 2023 - 00427.3002806535 - 2023 - 00428.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that the patient underwent a uni knee arthroplasty and approximately 7 months later a revision surgery was performed due to the failed implant on tibial side.There was a visible cyst about the size of a marble toward to posterior aspect of the tibia that was mostly contained.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned products identified the lot numbers of the tray and bearing match what is listed in the sub form of the complaint.All 3 items show sign of use with gouges and scratches on the bearing, scratches on the polished surface of the tray and the femoral component.The tray and femoral component also have blue residue (cement) attached as well.A review of the device manufacturing records confirmed no abnormalities or deviations.Devices are used for treatment.Medical records were not provided.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG LT SM SIZE 4 PMA
Type of Device
OXFORD HXLPE BEARINGS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18343140
MDR Text Key330660001
Report Number3002806535-2023-00429
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279785872
UDI-Public(01)05019279785872(17)280128(10)65677041
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number159541
Device Lot Number65677041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/26/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received01/30/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
-
-