Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).G2 - foreign: poland.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
Event Description
|
It was reported that a trial knee bearing fractured during a procedure.No piece fell or was left in the patient.;the damaged trial bearing was returned to the distributor warehouse.This event is related to a malfunction that could potentially lead to a sterility issue/serious injury.However, no patient harm or further outcome was reported.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).Visual examination of the returned product identified that the part was returned in 2 pieces.Minor scratches noticed on the part.No other damage noticed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device is used for treatment.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|