Model Number 865350
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Event Description
The customer reported that the speaker is not functioning.No sound is coming from the device at all.It is unknown if the device was in use at time of event, and there was no adverse event reported.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
Philips received a complaint on the mx40 1.4 ghz smart hopping device, indicating that the speaker is not functioning.No sound is coming from the device at all.The device was sent to philips authorized repair facility (rft) for bench for evaluation.Results of functional testing indicate the speaker had no sound in the mt56060 tool.They were unable to test speaker in the unit and speaker was defective.Based on the information available and the testing conducted, the cause of the reported problem was a defective speaker.The reported problem was confirmed.The customer was provided a replacement device to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18344506
MDR Text Key330780915
Report Number1218950-2023-00943
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received12/29/2023
Supplement Dates FDA Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-